CALCulate<1|2>:MARKer<1...4>:X:SSIZe
This command defines the step size of the rotary knob for marker or delta marker value
changes. It takes only effect in manual operation. It is available for all base unit measurements
with the exception of statistics.
The numeric suffixes <1|2> and <1...4> are irrelevant for this command.
This command is available from firmware version 1.60.
Parameter
STANdard measurement point step size
POINts measured value step size (number of measured values is defined via the
[SENSe<1|2>:]SWEep:POINts command)
Example
CALC:MARK:X:SSIZ POIN
Sets the measured value step size.
Characteristics
RST value: STANdard
SCPI: device–specific
Mode
A
CALCulate<1|2>:MARKer<1...4>:Y?
This command queries the measured value of the selected marker. The corresponding marker
is activated before or switched to marker mode, if necessary.
To obtain a correct query result, a complete sweep with synchronization to the sweep end must
be performed between the activation of the marker and the query of the Y value. This is only
possible in single sweep mode.
The query result is output in the unit determined with CALCulate<1|2>:UNIT:POWer.
In the default setting, the output is made depending on the unit determined with
CALCulate<1|2>:UNIT:POWer; only with linear level scaling is the output in %.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
Example
INIT:CONT OFF
Switches to single sweep mode.
CALC:MARK2 ON
Switches marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:Y?
Outputs the measured value of marker 2.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, BT, CATV, CDMA, EVDO, SPECM, WCDMA