CALCulate Subsystem R&S FSL
1300.2519.12 6.82 E-11
CALCulate:MARKer:FUNCtion:HARMonics Subsystem
The CALCulate:MARKer:FUNCtion:POWER subsystem contains the commands to define the settings
for harmonics measurement.
This subsystem is available from firmware version 1.10.
Commands of the CALCulate:MARKer:FUNCtion:HARMonics Subsystem
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics[:STATe]
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:BANDwidth:AUTO
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:DISTortion?
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:LIST?
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:NHARmonics
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:PRESet
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics[:STATe]
This command activates/deactivates the harmonic distortion measurement.
If the measurement is started in span > 0, the last span defines the search range for the first
harmonic. The level is determined for the first harmonic.
If the measurement is started in zero span, center frequency and level are used unchanged.
The numeric suffix <1|2> and <1...4> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
ON | OFF
Example
CALC:MARK:FUNC:HARM ON
Activates the harmonic distortion measurement.
Characteristics
RST value: OFF
SCPI: device–specific
Mode
A–F, A–T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:HARMonics:BANDwidth:AUTO
This command defines the resolution bandwidth of the harmonic in respect to the bandwidth of
the first harmonic. For details refer to chaper "Instrument Functions", section "Harmonic RBW
Auto".
The numeric suffix <1|2> are irrelevant for this command.
This command is available from firmware version 1.10.
Parameter
OFF identical
ON a multiple