R&S FSL CALCulate Subsystem
1300.2519.12 6.115 E-11
CALCulate:STATistics:SCALe:X:RANGe
This command defines the level range for the x–axis of the measurement diagram. The setting
is identical to the level range setting defined with the
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe] command.
Parameter
10dB to 200dB
Example
CALC:STAT:SCAL:X:RANG 20dB
Characteristics
*RST value: 100dB
SCPI: device–specific
Mode
A, CATV, CDMA, EVDO
CALCulate:STATistics:SCALe:X:RLEVel
This command defines the reference level for the x–axis of the measurement diagram. The
setting is identical to the reference level setting using the
DISPlay[:WINDow<1|2>]:TRACe<1...6>:Y[:SCALe]:RLEVel command.
With the reference level offset <> 0 the indicated value range of the reference level is modified
by the offset.
The unit depends on the setting performed with CALCulate<1|2>:UNIT:POWer.
Parameter
–120dBm to 20dBm
Example
CALC:STAT:SCAL:X:RLEV –60dBm
Characteristics
*RST value: –20dBm
SCPI: device–specific
Mode
A, CATV, CDMA, EVDO
CALCulate:STATistics:SCALe:Y:LOWer
This command defines the lower limit for the y–axis of the diagram in statistical measurements.
Since probabilities are specified on the y–axis, the entered numeric values are dimensionless.
Parameter
1E–9 to 0.1
Example
CALC:STAT:SCAL:Y:LOW 0.001
Characteristics
*RST value: 1E–6
SCPI: device–specific
Mode
A, CATV, CDMA, EVDO