Section 7. Installation
Table 36. PRTCalc() Type-Code-2 Sensor
US Industrial Standard, alpha = 0.00392 (Reference: Logan Enterprises)
Constant Coefficient
a 3.9786300E-03
d -2.3452400E-06
e 1.8174740E-05
f -1.1726200E-06
g 1.7043690E+00
h -2.7795010E+00
i 8.8078440E+00
j 2.5129740E+02
Table 37. PRTCalc() Type-Code-3 Sensor
US Industrial Standard, alpha = 0.00391 (Reference: OMIL R84 (2003))
Constant Coefficient
a 3.9690000E-03
d -2.3364000E-06
e 1.8089360E-05
f -1.1682000E-06
g 1.7010560E+00
h -2.6953500E+00
i 8.8564290E+00
j 2.5190880E+02
Table 38. PRTCalc() Type-Code-4 Sensor
Old Japanese Standard, alpha = 0.003916 (Reference: JIS C 1604:1981, National Instruments)
Constant Coefficient
a 3.9739000E-03
d -2.3480000E-06
e 1.8139880E-05
f -1.1740000E-06
g 1.7297410E+00
h -2.8905090E+00
i 8.8326690E+00
j 2.5159480E+02