Section 10. Troubleshooting
messages may not be obvious because the display is limited. Much of this
information is more easily accessed through the datalogger support software
(p. 95)
station status report. The message reports the following:
program compiled OK
warnings about possible problems
run-time errors
variables that caused out-of-bounds conditions
watchdog information
memory errors
Warning messages are posted by the CRBasic compiler to advise that some
expected feature may not work. Warnings are different from error messages in
that the program will still operate when a warning condition is identified.
A rare error is indicated by mem3 fail type messages. These messages can be
caused by random internal memory corruption. When seen on a regular basis with
a given program, an operating system error is indicated. Mem3 fail messages are
not caused by user error, and only rarely by a hardware fault. Report any
occurrence of this error to a Campbell Scientific application engineer, especially if
the problem is reproducible. Any program generating these errors is unlikely to be
running correctly.
Examples of some of the more common warning messages are listed in table
Warning Message Examples
(p. 486).
Table 130. Warning Message Examples
Message Meaning
CPU:DEFAULT.CR1 -- Compiled in
PipelineMode.
Error(s) in CPU:NewProg.CR1:
line 13: Undeclared variable Battvolt.
A new program sent to the datalogger failed to
compile, and the datalogger reverted to running
DEFAULT.cr1.
Warning: Cannot open include file CPU:
Filename.cr1
The filename in the Include instruction does not
match any file found on the specified drive.
Since it was not found, the portion of code
referenced by Include will not be executed.
Warning: Cannot open voice.txt
voice.txt, a file required for use with a COM310
voice phone modem, was not found on the CPU:
drive.
Warning: COM310 word list cannot be a
variable.
The Phrases parameter of the VoicePhrases()
instruction was assigned a variable name instead
of the required string of comma-separated words
from the Voice.TXT file.
Warning: Compact Flash Module not
detected: CardOut not used.
CardOut() instructions in the program will be
ignored because no CompactFlash (CF) card was
detected when the program compiled.
Warning: EndIf never reached at runtime.
Program will never execute the EndIf
instruction. In this case, the cause is a Scan()
with a Count parameter of 0, which creates an
infinite loop within the program logic.
486

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Campbell CR1000 and is the answer not in the manual?

Campbell CR1000 Specifications

General IconGeneral
Power Requirements9.6 to 16 VDC
Data Logger Operating SystemCRBasic
Memory4 MB
Analog Inputs16 single-ended or 8 differential
Analog Input Resolution13 bits
Analog Input Range+/-2.5V
Digital I/O8
Pulse Counters2
Communication PortsRS-232
Operating Temperature-25°C to +50°C
Pulse Channels4