EasyManua.ls Logo

Campbell CR1000

Campbell CR1000
678 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix B. Status, Settings, and Data Table Information (Status/Settings/DTI)
Battery (p. 473).
Low12VCount
Station Status: Number of times voltage has
dropped below 12V
Keyboard: Status Table ≈ line 21
CRBasic: variable = status.keyword
Increments by 1 each time the primary CR1000
supply voltage drops below ≈9.0. Updated with each
Status table update. The following applies to the
CR800 and CR1000 dataloggers. A variation may
apply to the C6 and CR3000: The 12 Vdc-low
comparator triggers at about 9.0 Vdc. The
minimum-specified input voltage of 9.6 Vdc will not
cause a 12 Vdc low condition, but a 12 Vdc low
condition will stop program execution before
measurements are compromised.
reset only
LONG
count
0
0 to 99
0 = reset
Low5VCount
Station Status: Number of times voltage has
dropped below 5V
Keyboard: Status Table ≈ line 22
CRBasic: variable = status.keyword
Number of occurrences of the 5 Vdc supply dropping
below a functional threshold.
reset only
LONG
count
0
0 to 99
0 = reset
Table 177. Status/Settings/DTI: M
Keyword
Alias,
Access,
Description
Read/Write,
DataType,
Units
Default
Value
Normal
Range
MaxBuffDepth
Keyboard: Status Table ≈ line 43
CRBasic: variable = status.keyword
Maximum number of buffers the CR1000 will use to
process lagged measurements.
read/write
LONG
no units
0
?
MaxPacketSize
Settings Editor: Max Packet Size
Keyboard: Settings (General) ≈ line 8
CRBasic: variable = settings.keyword;
SetSettings()
Maximum number of bytes per data collection
packet.
read/write
LONG
bytes
1000
?
MaxProcTime
Keyboard: Status Table ≈ line 41
CRBasic: variable = status.keyword
Maximum time required to run through processing
for the current scan. Value is reset when the scan
exits. Calculated on-the-fly. See MeasureTime
(p.
621),
ProcessTime (p. 625), SkippedScan (p. 628),
MaxProcTime
(p. 620).
read/write
LONG
μs
n/a 0 = reset
MaxSlowProcTime()
Keyboard: Status Table ≈ line 49
CRBasic: variable = status.keyword
Maximum time required to process a SlowSequence
scan in the CRBasic program. 0 until a scan runs.
See LastSlowScan
(p. 619), SkippedSlowScan (p. 628),
SlowProcTime
(p. 629).
read/write
LONG array
μs
0 until scan
0 = rest
620

Table of Contents

Related product manuals