EasyManuals Logo

Campbell CR1000 User Manual

Campbell CR1000
678 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #274 background imageLoading...
Page #274 background image
Section 7. Installation
The SDIRecorder() aCv command facilitates using the SDI-12 standard Start
Concurrent command (aCv!) without the back-to-back measurement sequence
normal to the CR1000 implementation of aCv!.
Consider an application wherein four SDI-12 temperature sensors need to be near-
simultaneously measured at a five minute interval within a program that scans
every five seconds. The sensors requires 95 seconds to respond with data after a
measurement request. Complicating the application is the need for minimum
power usage, so the sensors must power down after each measurement.
This application provides a focal point for considering several measurement
strategies. The simplest measurement is to issue a M! measurement command to
each sensor as shown in the following CRBasic example:
Public BatteryVolt
Public Temp1, Temp2, Temp3, Temp4
BeginProg
Scan(5,Sec,0,0)
'Non-SDI-12 measurements here
SDI12Recorder(Temp1,1,0,"M!",1.0,0)
SDI12Recorder(Temp2,1,1,"M!",1.0,0)
SDI12Recorder(Temp3,1,2,"M!",1.0,0)
SDI12Recorder(Temp4,1,3,"M!",1.0,0)
NextScan
EndProg
However, the code sequence has three problems:
1. It does not allow measurement of non-SDI-12 sensors at the required frequency
because the SDI12Recorder() instruction takes too much time.
2. It does not achieve required five-minute sample rate because each
SDI12Recorder() instruction will take about 95 seconds to complete before
the next SDI12Recorder() instruction begins, resulting is a real scan rate of
about 6.5 minutes.
3. There is a 95 s time skew between each sensor measurement.
Problem 1 can be remedied by putting the SDI-12 measurements in a
SlowSequence scan. Doing so allows the SDI-12 routine to run its course
without affecting measurement of other sensors, as follows:
Public BatteryVolt
Public Temp(4)
BeginProg
Scan(5,Sec,0,0)
'Non-SDI-12 measurements here
NextScan
SlowSequence
Scan(5,Min,0,0)
SDI12Recorder(Temp(1),1,0,"M!",1.0,0)
SDI12Recorder(Temp(2),1,1,"M!",1.0,0)
SDI12Recorder(Temp(3),1,2,"M!",1.0,0)
274

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Campbell CR1000 and is the answer not in the manual?

Campbell CR1000 Specifications

General IconGeneral
Power Requirements9.6 to 16 VDC
Data Logger Operating SystemCRBasic
Memory4 MB
Analog Inputs16 single-ended or 8 differential
Analog Input Resolution13 bits
Analog Input Range+/-2.5V
Digital I/O8
Pulse Counters2
Communication PortsRS-232
Operating Temperature-25°C to +50°C
Pulse Channels4

Related product manuals