EasyManuals Logo
Home>Campbell>Measuring Instruments>CR1000

Campbell CR1000 User Manual

Campbell CR1000
678 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #345 background imageLoading...
Page #345 background image
Section 8. Operation
worst-case is (91 segments) • (4 s / segment) = 364 s per complete self-calibration.
During instrument power-up, the CR1000 computes calibration coefficients by
averaging ten complete sets of self-calibration measurements. After power up,
newly determined G and B values are low-pass filtered as follows:
Next_Value = (1/5) • (new value) + (4/5) • (old value)
This results in the following settling percentages:
20% for 1 new value,
49% for 3 new values
67% for 5 new values
89% for 10 new values
96% for 14 new values
If this rate of update is too slow, the Calibrate() instruction can be used. The
Calibrate() instruction computes the necessary G and B values every scan
without any low-pass filtering.
For a VoltSe() instruction, B is determined as part of self-calibration only if the
parameter MeasOff = 0. An exception is B for VoltSe() on the ±2500 input range
with a 250 µs integration, which is always determined in self-calibration for use
internally. For a VoltDiff() instruction, B is determined as part of self-calibration
only if the parameter RevDiff = 0.
VoltSe() and VoltDiff() instructions, on a given input range with the same
integration durations, use the same G values but different B values. The six input-
voltage ranges (±5000 mV, ±2500 mV, ±250 mV, and ±25 mV), in combination
with the three most common integration durations (250 µs, 50 Hz half-cycle, and
60 Hz half-cycle) result in a maximum of 18 different gains (G), and 18 offsets for
VoltSe() measurements (B), and 18 offsets for VoltDiff() measurements (B) to be
determined during CR1000 self-calibration (maximum of 54 values). These
values can be viewed in the Status table, with entries identified as listed in table
Status Table Calibration Entries
(p. 346).
Automatic self-calibration can be overridden with the Calibrate() instruction,
which forces a calibration for each execution, and does not employ low-pass
filtering on the newly determined G and B values. The Calibrate() instruction
has two parameters: CalRange and Dest. CalRange determines whether to
calibrate only the necessary input ranges for a given CRBasic program (CalRange
= 0) or to calibrate all input ranges (CalRange 0). The Dest parameter should
be of sufficient dimension for all returned G and B values, which is a minimum of
two for the automatic self-calibration of VoltSE() including B (offset) for the
±2500 mV input range with first 250 µs integration, and a maximum of 54 for all
input-voltage ranges used and possible integration durations.
An example use of the Calibrate() instruction programmed to calibrate all input
ranges is given in the following CRBasic code snip:
'Calibrate(Dest,Range)
Calibrate(cal(1),true)
where Dest is an array of 54 variables, and Range 0 to calibrate all input ranges.
Results of this command are listed in the table Calibrate() Instruction Results
(p.
347).
345

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Campbell CR1000 and is the answer not in the manual?

Campbell CR1000 Specifications

General IconGeneral
BrandCampbell
ModelCR1000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals