40500
40500
Attention message: Processor
No. 00503[N] [SM]
Empty the Effluent Tank.
Attention message release
YES or automatic release
• This attention message will be automatically closed when the effluent tank is emptied.
Condition
Check Point
Diagnosis
$ Symptoms due to connection failure of wiring/symptoms resulted from blowout of fuse
NOTE
• To access the wiring diagrams, click Wiring diagrams in Bookmarks.
Suffix number Condition
00001 The CD effluent float switch turns on.
00002 The BF effluent float switch turns on.
00004 The STB effluent float switch turns on.
00008 P2 + PS Effluent Float Switch turns on.
1 Check the solution level of the effluent tank.
2 Check if the float of the effluent float switch moves up and down smoothly.
Failed part Manual No.
Effluent float switch
☞63290
Processor control PCB ☞66200
Processor I/O PCB 1 ☞66210
Processor power supply 2
☞67400
Symptoms due to the wiring connection failure Symptoms of fuse blowout
− ☞4203 −