Processor Settings
1/3
3. Mode
33003
33003
Pump Output Amount Setting[F]
Description for each specification
In this section, the location of information varies depending on the specification. See below for the description for each specification.
• For [N] and [SM] specifications, see Pump Output Amount Setting [N] [SM].
!
!!
! Display
Bringing up the display
Maintenance: 2260ExtensionProcessor SettingsPump Output Amount Setting
!
!!
! Explanation
!
!!
! Pump Output Amount Setting (Input range: 0.1 to 200.0 ml)
By Pump Output Amount Measurement via F: Functions, output amount of each replenisher pump is measured, and the measured
values are entered.
!
!!
! F: Functions
!
!!
! Pump Output Amount Measurement
Use this function to measure the output amount of each pump.
For the operation procedures, see Operator's Manual - Additional Operations - Measuring the pump output amount.
!
!!
! Mixing Replenisher
There are two functions in Manual Open: creating replenishment solution and resetting the remaining amount of the replenishment
solution.
Do not return the replenishment solution used for measuring the output amount to the replenishment tank because this amount is
subtracted from the total replenishment solution amounts.
When restarting the process without draining the replenishment solution from P1R/P2RA/P2RB after taking countermeasures
against replenisher section errors,
No. 00912[F] is shown if adding water to the replenishment tank and performing Mixing
Replenisher with turning on all the (upper) and (lower) replenishment solution level sensors (ON: the state of solution filled).
At this time, the remaining amount of replenishment solution is reset.
No. 00913[F] occurs if performing Mixing Replenisher with # # # (lower) replenishment solution level sensor on (EXCEPT FOR
the case that all the (upper) and (lower) replenishment solution level sensors of P1R/P2RA/P2RB are on, which is the state of
solution filled).
The same process can be performed as Processor Standard SettingProcessor Set Up ModeMixing Replenisher.
M5467-00