Frontier LP7700 - No. 05916[F] Solution Level of the Processing Tank Goes below the Processable Level. P; No. 05917[F] Solution Level of the Processing Tank Goes below the Processable Level. P; No. 05918[F] Solution Level of the Processing Tank Goes below the Processable Level. Ps; No. 05919[F] Solution Level of the Processing Tank Goes below the Processable Level. Ps

Frontier LP7700
904 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. Troubleshooting
45900
12/16
45900
Error message: F replenishment
No. 05916[F]
Solution level of the processing tank goes below the processable level. P1
No. 05917[F]
Solution level of the processing tank goes below the processable level. P2
No. 05918[F]
Solution level of the processing tank goes below the processable level. PS1
No. 05919[F]
Solution level of the processing tank goes below the processable level. PS2
No. 05920[F]
Solution level of the processing tank goes below the processable level. PS3
No. 05921[F]
Solution level of the processing tank goes below the processable level. PS4
Error message release
YES
Condition
NOTE
If the processing solution level detector of the processing solution float switch turns off, all the processing solution heaters and
circulation pumps turn off.
Check Point
*1. If Evaporation Correction is set to OFF, water is not replenished automatically although # # # processable level decreases.
Diagnosis
Suffix number Condition
00000 # # #When the processable level decreases and addition of specific amount water is carried, # # #the
processable level detector keeps off.
When the product starts up with program timer and filling up the moisture which evaporated is carried out,
# # # the processable level detector keeps off.
• If Pump Output Amount Measurement of Pump Output Amount Setting and Measure Auto Cleaning Water Output
Amount of Auto Cleaning Water Output Amount Setting are not performed at system installation, the above errors may occur.
The causes are shown below.
1. Even if specified amount of water is supplied, # # # processable level does not turn on.
2. Even if each replenishment pump operates for a specified time, specified amount of solution is not replenished.
Countermeasure
1. Perform Taking countermeasures after the errors No.05916 to No.05921 occurs.
4630
Check if the processing solution is leaking from the tank. -
Check if the float of the processing solution float switch moves up and down smoothly. -
Check the status of the float switches of each processing solution via Input Check. 35220
Check each replenishment pump output amount again.
33003
Check the output amounts of auto cleaning pump + each automated cleaning valve again. 33004
Select Operator SelectionsProcessor, and change the Evaporation Correction setting from OFF to
ON.
*1
-
When the level is incorrectly detected Manual No.
Processing solution float switch
63280