4. Troubleshooting
45530
3/14
45530
Error: Processor 2
No. 05533
The Dryer Safety Thermostat has activated.
Error message release
YES
Condition
NOTE
The dryer safety thermostat is activated at the following temperatures.
When the dryer safety thermostat goes OFF, the dryer heater and dryer fan go OFF.
Check Point
Diagnosis
$ Symptoms due to connection failure of wiring/symptoms resulted from blowout of fuse
NOTE
To access the wiring diagrams, click Wiring diagrams in Bookmarks.
Suffix number Condition
00000 The dryer safety thermostat turns off.
Status Temperature (°C)
OPEN(OFF) 90±2.8
CLOSE(ON) 70±4.4
1 Check if the dryer fan is rotating.
When the dryer fan rotates incorrectly Manual No.
Dryer fan
63260
When temperature is incorrectly adjusted Manual No.
Processor control PCB 66200
When the safety thermostat is defective Manual No.
Dryer safety thermostat 63260
Processor I/O PCB 1 66210
Processor relay PCB
66220
Processor control PCB
66200
When temperature is incorrectly detected Manual No.
Dryer thermosensor
63260
Processor control PCB 66200
Processor relay PCB 66220
Processor power supply 2
67400
Symptoms due to the wiring connection failure Symptoms of fuse blowout
4203 4252