LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
BLACK-OUT
Black-out Mains voltage < 160V
Sudden power failure (black-out)
Wait for electricity to return, open the door and follow the
instructions on the LCD.
Reset the alarm and repeat the sterilization cycle.
Accidentally turning-off the main switch
Switch on the equipment, open the door and follow the
instructions on the LCD.
Reset the alarm and repeat the sterilization cycle.
Accidentally pulling the plug out of the
socket.
Reconnect the plug, open the door and follow the instructions
on the LCD.
Reset the alarm and repeat the sterilization cycle.
Mains fuses blown.
Replace the burned fuse (16A), switch on the equipment, open
the door and follow the instructions on the LCD.
Reset the alarm and repeat the sterilization cycle.
Pay attention during the occurring of the
alarm and find the component causing the
fault
Use the repair layout for the involved component
The alarm occurs each time the equipment
is turned on
Check that the operator resets correctly the alarm.
Explain the correct procedure to be used.
Check the main switch.
Replace the main switch - see card Gr1-17
Check for possible water or steam on electric components.
Check the steam generator cartridges.
Replace the steam generator cartridge – see card
Gr1-19 .
The alarm occurs when opening the door
and starts the report printout.
Check and replace the fuse on the power supply printer board.
Replace the filter/PS printer board – see card Gr1-1.