Rev.1
Page 1-
13
500
570
550
GENERAL INSTALLATION PRECAUTIONS
Obey the following warnings for the correct operation of the device and/or to avoid risky situations:
– Install the sterilizer on a flat surface
; if necessary, adjust the leveling feet to compensate for an irregular surface and to slightly tilt lower the front part of the
sterilizer.
Make sure that the support surface is strong enough
to support the equipment weight (about 60 kg);
– Leave adequate space for ventilation (at least 10 cm on each side) all around the sterilizer, especially in back.
If the device is built-in to a cabinet, be sure to respect the warnings in the preceding paragraph, avoiding an obstructions to the air intake;
– Do not install the sterilizer near tubs, sinks or similar places, to avoid contact with water or liquids. This could cause short circuits and/or potentially
dangerous situations for the operator;
– Do not install the sterilizer in a place that is excessively humid
or poorly ventilated;
– Do not install the machine were there is gas
or inflammable and/or explosive vapors;
– Install the device so that the power cord is not
bent or crushed. It must run freely all the way to the socket.
– Install the device that any external fill/drain tubing is not
bent or crushed. They must run freely to the drain tank.
COMPARTMENT DIMENSIONS FOR BUILT-IN INSTALLATIONS
When installing the sterilizer inside a cabinet, you must provide adequate space all around the device (>10cm, specially in the rear part) to
provide effective ventilation as well as a large enough opening in the back that, in addition to allowing the passage of the power cord will also
provide an adequate air flow and the consequent optimum cooling of the heat exchanger.
It is indispensable that the built-in compartment have the following minimum dimensions
shown in the figure:
C
OMPARTMENT DIMENSIONS LESS THAN THOSE SHOWN MAY COMPROMISE THE CORRECT CIRCULATION
OF AIR AROUND THE DEVICE AND MAY NOT PROVIDE ADEQUATE COOLING
, WITH THE CONSEQUENT
DETERIORATION OF PERFORMANCE AND
/OR POSSIBLE DAMAGE.
e If the main switch is inaccessible when installed in the compartment, use an electric plug that
incorporates an on/off switch.
DO NOT REMOVE THE UPPER COVER OR ANY OTHER EXTERNAL PART. WHEN INSTALLED IN THE
COMPARTMENT
, THE DEVICE MUST BE COMPLETE WITH ALL ITS PARTS.
e Please refer to Technical Characteristics for complete technical data.