Rev.1
Page 1-
26
Setting the Time
(TIME SETTING on the BASIC menu)
When TIME SETTING is selected with the ↵ key, you will see:
hh:mm:ss
+/-to set
↵ to enter
to exit
Proceed as follows:
– The hours flash: set the current hour with the + and - keys. Confirm with ↵.
– The minutes flash: set the current value with the + and - keys. Confirm with ↵.
The time is stored. Once the last confirmation is given, you return to the second-level menu.
Setting the Password
(PASSWORD on the BASIC menu)
When PASSWORD is selected with the ↵ key, you will see this menu:
→ DISABLED +
ANY POWER ON ↑
ANY CYCLE START ↓
EXIT -
Select DISABLED to use the device freely, without any limitation on operator access.
Select ANY POWER-ON to protect the machine with a password at the time it is turned-on (power-on from the main switch).
This makes sure that the machine can only be powered-on by authorized personnel, but afterwards it can be used by others without limitation.
Select ANY CYCLE START to protect the autoclave with a password to be entered both at power-on and at the start of every sterilization
program.
Only authorized personnel will be able to use it.