Rev. 1
Page 3-
56
A 260
LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
PROCESS TIMEOUT
Depressurization not
performed within the
preset time
Problem on switching the
valves during the steam
discharge
EV3 interrupted
Replace the EV3 shaft - see card Gr2-3
EV3 Shaft blocked
EV3 dirty
Clean or replace the shaft - see card Gr2-3
Replace the wiring
Replace the fuse - see card Gr1-16
EV3 not powered
Replace the CPU board - see card Gr1-1
Check the arrangement of the sterilizer
(slope)
Change as necessary (adjust the feet for the proper
slope)