LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
Replace the water pump - see card Gr3-1
ATM1 TIMEOUT
1
st
pressure pulse
(from -.88 to 0.00) not
performed within the
preset time
1
st
pressure pulse not
performed within 3’ timeout
The water pump of the steam generator
does not work properly
Replace the power board - see card Gr1-1
Clean the valve
Incorrect operation of the EV6 valve
Replace EV6 valve - see card Gr3-1
Fault in the pipe connecting the steam
generator
Replace the pipe
Sterilization chamber overloaded
Advice the operator for properly loading of the sterilization
chamber.
Restore the safety thermostat and repeat the sterilization cycle
Intervention of the steam generator safety
thermostat
Replace the safety thermostat - see card Gr1-12
Replace the lower section of the steam generator
- see card Gr4-6
Steam generator clogged
Different distilled water must be used
Fault in the board
Replace the board - see card Gr1-1
CPU board uncalibrated
Send the sterilizer to the Mocom Service department
Check the continuity of the steam generator
cartridge
Replace the failed cartridge