Rev. 1
Page 3-
19
E 031
LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
EXHAUST MAX
Water at maximum
level in the recovery
tank .
The cycle does not start
Cycle started with water level over the
maximum threshold.
Empty completely the recovery tank.
Check for possible failure of the float wiring.
Restore the float wiring
Failure of the float.
Replace the float - see card Gr4-2
Check the wiring of the water level sensor
for the external used water tank.
Disconnect the wiring on the CPU board
Check the water level sensor connection on
the CPU board.
Restore the connection.
Replace the float - see card Gr4-2
Replace the CPU board – see card Gr1-1
Sterilizer configuration set on external water
drain.
Restore the connection of the external water level
sensor.
Empty the external used water tank.
Replace the external water level sensor.
Replace the wiring of the external water level sensor.
Wrong water drain sterilizer configuration.
Set the proper water drain option in the SETUP
menu.