ATTACHMENT A -
ADJUST DOOR POSITIONER
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card Gr7-1);
2. Open and close the door to check its operation;
3. In case the door does not remain close (no click at the stroke end), loosen
the nut (1) and turn clockwise the positioner (2);
4. In case of resistance on closing/opening the door, loosen the nut and turn
anti-clockwise the positioner;
5. Tighten the nut;
6. Perform a sterilization cycle.