Card Gr6-
3
DOOR BUSH
490#00050000
DOOR FORK AND PIN
C#0BP0500000
DOOR NITRIDED FORK
C#0BP060000Z
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Open the door;
Bushes / Pin
2. Remove the Seiger (1) from the bottom of the pin;
3. Withdraw the pin (2);
4. Remove the bushes (3);
Fork
5. Remove the door cover (4) (see card Gr7-2);
6. Remove screws and washers (5) and remove the fork (6);
7. Mount the new part, assembly all items proceeding in reverse order as
above;
8. Perform a sterilization cycle.
WARNING
The pin must result with the lowered side on top of the fork.