H 993
LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
OVERHEATING
PT3
PT3 overheating PT3 detects a temperature
value greater than 160°C
Fault in the board
Replace the CPU board- see card Gr1-1
Fault in the pressure transducer
Replace the pressure transducer - see card Gr1-4
Stored calibration values changed
Send the sterilizer to the Mocom Service department