LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
OVER PRESSURE
Overpressure in the
sterilization chamber
Pressure value greater than
2.32 bar
Fault in the board
Replace the CPU board - see card Gr1-1
Fault in the pressure transducer
Replace the pressure transducer - see card Gr1-4
Stored calibration values changed
Send the sterilizer to the Mocom Service department