ACTIVATING CONFIGURATION OPTIONS – ADVANCED MENU
Setting the sterilization programs
(PROGRAMMES on the ADVANCED menu)
The program setting and their storing in four pre-set positions is achieved in various steps using several menus in sequence.
Each pre-set position can be associated to a standard or user configurable cycle (CUSTOM).
Let's look at the two cases separately.
To associate a standard program and define several of its parameters, proceed as follows:
1. Select PROGRAMS using the ↵ key; the following menu appears:
→ 1st PRESET +
2nd PRESET ↑
3rd PRESET ↓
4th PRESET -
EXIT
Define the position (1, 2, 3 or 4) to which the sterilization program will be associated using the + and - keys. Confirm with the ↵ key.
2. From here, you enter the list of available cycles:
→ 134°C HOLLOW +
121°C HOLLOW ↑
134°C SOLID ↓
121°C SOLID -
134°C EMERGENCY
134°C WRAPPED
121°C WRAPPED
134°C POROUS
121°C PROROUS
134°C PRION
XXX°C CUSTOM
EXIT