EasyManua.ls Logo

Mocom millennium B - H 400

Mocom millennium B
285 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rev. 1
Page 3-
59
H 400
LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
P/T
PROBLEM
Ratio P
conv
/T not
correctly balanced
(P
conv
greater than T)
during the process
phase
The difference value between
P
conv
and T is greater of 2°C
Problem in the water pump of the steam
generator
Replace the water pump - see card Gr3-1
Clean the valve and repeat the sterilizaztion cycle
EV6 does not open correctly
Replace EV6 - see card Gr3-1
Identify the valve causing the problem and clean it
Steam leakage from a valve
Replace the valve causing the problem
- see cards Gr2
The discharge valve does not open correctly
Replace the discharge valve
Fault in the board
Replace the CPU board - see card Gr1-1
Stored calibration values changed
Send the sterilizer to the Mocom Service department
Steam leakage from the door gasket
Replace the door gasket - see card Gr6-1
Check the quality of the distilled water
Clean the hydraulic circuit
Replace the damaged components
Advice the user
Check the PCB filter board
replace the board – see card Gr1-1

Table of Contents