MAINTENANCE
GENERAL
For better quality maintenance, supplement ordinary checks with regular periodic examinations by the service department (see Appendix Z).
It is also fundamental to perform a periodic sterilizer validation
, i.e., a check of the thermodynamic parameters of the process, comparing them
with the reference values provided with suitably calibrated instruments.
e Refer to the paragraph Periodic sterilizer validation.
The ordinary maintenance described below consists in easy manual operations and preventive interventions involving simple instruments.
I
N THE EVENT OF THE REPLACEMENT OF THE DEVICE'S COMPONENTS OR PARTS, REQUEST AND/OR USE ORIGINAL REPLACEMENT PARTS
ONLY.
ORDINARY MAINTENANCE PROGRAM
Maintenance schedule
The table summarizes the maintenance required to keep the sterilizer operating at peak efficiency. In the case of very intense use, we
recommend shortening
maintenance intervals:
DAILY
Clean the gasket and the porthole
Clean external surfaces
Cleaning of the gasket and do
Cleaning of the external
WEEKLY
Clean the sterilization chamber and relative accessories
Disinfect external surfaces
Cleaning of the sterilization chamb
Disinfection of the extern
MONTLY
Clean the internal (and external - if installed) distilled water tank
Safety valve maintenance
Clean (or replace) the water drain filter
Lubrication of the door locking mechanis
Maintenance of the safety valve
Cleaning (or replacement) of the draining
Sterilization of the bacteriologic filter
EVERY 3/6 MONTHS
(depending on the frequency of
use)
Replace bacteriological filter
Replacement of the bacteriologic filter
ANNUALLY
Validate sterilizer
Validation of the sterilizer