EasyManua.ls Logo

Mocom millennium B - HAZARD ALARM LIST - “H” CODES

Mocom millennium B
285 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Rev. 1
Page 3-
10
HAZARD ALARM LIST - “H” CODES
CODE ALARM DESCRIPTION LCD INDICATION RESET MODE
H 150
MPX pressure sensor broken MPX BROKEN
H 160
MPX pressure sensor short-circuited/not
connected
MPX SHORTCIRCUIT
Turning off the
equipment
H 400
Ratio P
conv
/T not balanced (P
conv
>T)
(Phase PROCESS)
P/T PROBLEM
H 401
Ratio T/P
conv
not balanced (T>P
conv
)
(Phase PROCESS)
T/P PROBLEM
H 402
Temperature above MAX limit
(Phase PROCESS)
T OVER LIMIT
H 403
Temperature below MIN limit
(Phase PROCESS)
T UNDER LIMIT
H 404
Temperature fluctuating over the limit
(Phase PROCESS)
PT1 FLUCTUATING
H 405
Pressure above MAX limit
(Phase PROCESS)
P OVER LIMIT
H 406
Pressure below MIN limit
(Phase PROCESS)
P UNDER LIMIT
H 410
Wrong maintenance time
(Phase PROCESS)
TIMER PROBLEM
H 990
Excessive pressure
(sterilization chamber, MPX)
OVERPRESSURE
H 991
Overheating
(sterilization chamber, PT1)
OVERHEATING PT1
H 992
Overheating
(steam generator, PT2)
OVERHEATING PT2
H 993
Overheating
(band heating element, PT3)
OVERHEATING PT3
Hold down
(> 3 seconds)

Table of Contents

Related product manuals