E 010
LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
DOOR OPEN
Door open
The door micro-switch is not
activated and remains in OFF
position
Door not
properly closed.
Explain the proper procedure on closing and opening the door.
The door does not close completely
Check the positioning pin.
- Adjust the positioner – see Attachment A
- Replace the positioner – see card Gr6-4
The door opens by oneself
- Adjust the positioner – see Attachment A
- Replace the positioner – see card Gr6-4
After the door gasket replacement, the door
rebounds.
Call the Service, communicate the serial number of the
sterilizer and follow the indications suggested.
- Replace the door gasket - see card Gr6-1
- Replace the parabola- see card Gr6-2
Door properly closed but the door icon
remains off.
Check the door microswitch.
Unlock the door microswitch.
Replace the door microswitch – see card Gr6-5.
Damage during the equipment transport
Replace the damaged part.