Rev. 1
Card Gr3-
2
AUTOMATIC WATER FILLING PUMP ASSEMBLY
A#0BP2870000
AUTOMATIC WATER FILLING PUMP
400#00200000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card Gr7-1 );
2. Remove the pipes and the wiring from the automatic water feeding pump;
3. Remove the screws fixing the water pump assembly to the rear frame;
4. Disassembly the pump as necessary (see figure);
5. Mount the new pump assembly, restore connections and assembly all items
proceeding in reverse order as above;
6. Perform an automatic water feeding;
7. Run a sterilization cycle.