Rev.1
Page 1-
14
ELECTRICAL CONNECTIONS
The sterilizer's must be connected to a socket of the electrical system of adequate capacity for the device's absorption and ground provided, in
conformity with current laws and/or standards.
The socket must be suitably protected by breakers having the following characteristics:
– Nominal current I
n
16 A
– Differential current I
∆n
0,03 A
e The manufacturer will not be liable for damages caused by installing the sterilizer on an inadequate electrical system and/or not
equipped with a ground.
If it is necessary to replace the plug on the power cord, use one with equal characteristics or, at any rate, adequate to the device's electrical
characteristics. The user is entirely responsible for the selection and replacement of the plug.
e Always connect the power cord directly to the socket. Do not use extension cords, adapters or other accessories.
CONNECTING THE DATA RECORDING MILLFLASH
The sterilizer can be connected to MILLFLASH allowing the recording of the cycle data in .txt format file and its management by PC.
The connectors of the service box are used for interfacing; refer to MILLFLASH Operating Manual for the installation instructions.
Differential switch
Mains switch
Mains socket
with earth pin