Thermodynamic equilibration
When the pressure and temperature values for the selected program have been reached, it is a good idea to wait a moment to allow the temperature in the chamber and the load
to stabilize (EQUILIBRATION). The liquid crystal display shows:
1 134°C POROUS
EQUILIBRATION
135.4 °C
2.15 bar 13:40
CHAMBER
Water
tank
DESCRIPTION
STATUSES
(P, t)
VP WP
EV1
(no)
EV2
(nc)
EV3
(nc)
EV4
EV5
(no)
EV6
(nc)
Fan
Equilibration
P > +2,15 bar x 15” (134°C)
P > +1,12 bar x 15” (121°C)
OFF ON ON OFF OFF OFF ON ON ON