Rev. 1
Card Gr4-
11
SAFETY VALVE ASSEMBLY
A#0BS0520000
SAFETY VALVE
470#00020000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card Gr7-1 );
2. Remove the sheathed pipe from the safety valve fitting;
3. Remove the safety valve assembly (1);
4. Remove the safety valve (2) from the assembly;
5. Mount the new valve, assembly all items and restore connections proceeding
in reverse order as above;
6. Run a sterilization cycle.