Clean sterilization chamber and accessories
Clean the sterilization chamber, support and trays (and internal surfaces in general) with a clean cotton cloth soaked in water and, possibly, the
addition of a small amount of neutral detergent. Carefully rinse with distilled water, taking care not to leave any type of residue in the chamber or
on accessories.
e Do not use sharp or pointed instruments to remove lime encrustation from the sterilization chamber. Whenever there are visible deposits,
immediately check the quality of the distilled water used (see Water supply characteristics).
Disinfect external surfaces
For the occasional disinfection of the external surfaces, you can use either denatured alcohol or detergents with a minimum percentage of
sodium hypochlorite (or equivalent).
Clean internal distilled water tank
1. Arrange an empty container on the floor near the sterilizer and put the free end of a tube into it.
2. Unscrew the plug (1) from the rear draining point and plug-in the other end of the tube.
3. Wait until the internal tank is completely emptied; close the drain point with the plug.
4. Prepare 4 liter of distilled water mixed with 10% of pure alcohol and fill the supplied standard container
5. Now fill completely the internal tank with this solution (see chapter ā€œFilling distilled waterā€ for the procedure) and allow the solution to sit for
30 minutes.
e Do not run any cycle during this period.
6. Drain again the internal tank and discard the solution. Close the drain point with the plug.
Clean external distilled water tank
1. Disconnect the external tank from the sterilizer; eventually recover the distilled water contained in it.
2. Fill the tank with a solution of distilled water and alcohol (10%)
3. Allow the solution to sit for 30 minutes.
4. Drain the tank and discard the solution.
5. Reconnect the tank to the sterilizer.