Rev. 1
Attach.
14
ATTACHMENT L -
RELEASE ALARM A022 - DOOR LOCKED
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card Gr7-1 );
2. Unplug jumper X21 of the CPU board;
3. Push the key Start and switch-on the unit;
4. LCD will show the message “SETUP”;
5. Press the Start key up to the message “SETUP COMPLETE”;
6. Wait for the end of the Self test, then switch-off the unit;
7. Restore the jumper X21 and switch on again the unit;
8. Perform a sterilization cycle.