Rev. 1
Page 2-
3
PROGRAM SUMMARY TABLE
NOMINAL VALUES
BASIC PROGRAM
PARAMETERS
STERILIZABLE MATERIAL
PROGRAM
DESCRIPTION
Temperature (°C)
Pressure (bar)
Holding time (min)
Cycle type
(EN 13060: 2004)
Pre-vacuum
(F=fractionated; S=single)
Standard drying
(L=long; S=short)
Total cycle time
(average load ÷ max load)
Average consumption
H
2
O (ml/cycle)
Average energy
consumption (kWh/cycle)
TYPE
MAX TOTAL MASS
(kg)
MAX MASS
PER TRAY (kg)
MAX MASS
PER ARTICLE (kg)
NOTES
Porous, unpackaged
material
1,00 0,30 0,30
Porous material in
single package
0,75 0,25 0,25
Porous material in
double package
0,60 0,20 0,20
Solid and hollow
material in single
package
3,00 1,00 0,25
134°C
POROUS
134
2,10
4
B
F
L
39÷44
525
0,8
Solid and hollow
instruments in
double package
1,50 0,50 0,25
Porous, unpackaged
material
1,00 0,30 0,30
Porous material in
single package
0,75 0,25 0,25
Porous material in
double package
0,60 0,20 0,20
Hollow instruments
in single package
3,00 1,00 0,25
134°C
PRION
134
2,10
>18
B
F
L
51÷56
550
0,9
Solid and hollow
instruments in
double package
1,50 0,50 0,25
Porous, unpackaged
material
1,00 0,30 0,30
Porous material in
single package
0,75 0,25 0,25
Porous material in
double package
0,60 0,20 0,20
Hollow instruments
in single package
3,00 1,00 0,25
121°C
POROUS
121
1,10
20
B
F
L
54÷59
550
0,8
Solid and hollow
instruments in
double package
1,50 0,50 0,25
For material and instruments in (single and double) packaging, we recommend
using the 3-tray configuration
134°C
HOLLOW
134
2,10
4
S
F
S
30÷33
525
0,7
Unpackaged hollow
instruments
6,00 1,20 0,50
NOMINAL VALUES
BASIC PARAMETERS OF THE
PROGRAM
STERILIZABLE MATERIAL
PROGRAM
DESCRIPTION
Temperature (°C)
Pressure (bar)
Holding time (min)
Cycle type
(EN 13060: 2004)
Pre-vacuum
(F=fractionated; S=single)
Standard drying
(L=long; S=short)
Total cycle time
(average load ÷ max load)
Average consumption
H
2
O (ml/cycle)
Average energy
consumption (kWh/cycle)
TYPE
MAX TOTAL MASS
(kg)
MAX MASS
PER TRAY (kg)
MAX MASS
PER ARTICLE (kg)
NOTES
121°C
HOLLOW
121
1,10
20
S
F
S
44÷47
550
0,7
Unpackaged
hollow
instruments
6,00 1,20 0,50
134°C
WRAPPED
134
2,10
4
S
S
L
33÷36
300
0,6
Unpackaged
hollow
instruments
3,00 1,00 0,25
121°C
WRAPPED
121
1,10
20
S
S
L
47÷50
325
0,6
Solid
instruments in
single package
3,00 1,00 0,25
We recommend
using the 3-tray
configuration
134°C
SOLID
134
2,10
4
N
S
S
25÷28
300
0,5
Solid
instruments in
single package
6,00 1,20 0,50
121°C
SOLID
121
1,10
20
N
S
S
39÷42
325
0,5
Unpackaged
solid instruments
6,00 1,20 0,50
134°C
EMERGENC
Y
134
2,10
3
N
S
Fast
16
300
0,45
Unpackaged
solid instruments
0,50 0,50 0,50
XXX°C
USER
134 or
121
2.10
or 1.10
> 4 or
> 20
n.d.
F/S
L/S
n.d.
n.d.
n.d.
Unpackaged
solid instruments
n.d. n.d. n.d.
Variable
paramet
.
HELIX/BD
TEST
134
2.10
3.5
-
F
S
22
-
-
Unpackaged
solid instruments
- - -
VACUUM
TEST
-
-0.80
-
-
-
-
22
-
-
Test device only
(no other load)
- - -