Rev. 1
Page 3-
8
ALARM LIST - “A” CODES
CODE ALARM DESCRIPTION LCD INDICATION RESET MODE
A 022
System door lock microswitches failed
(OFF-OFF)
LOCKING PROBLEM
A 023
System door lock microswitches failed
(ON-ON)
LOCKING PROBLEM
A 024
System door lock microswitches failed
(ON-OFF)
LOCKING PROBLEM
A 032
Sensor-level problem LEVEL PROBLEM
A 040
Failure to fill the tank
(automatic filling)
FILLING PROBLEM
A 101
PT1 broken
(sterilization chamber)
PTC BROKEN
A 102
PT2 broken
(steam generator)
PTC BROKEN
A 103
PT3 broken
(heating element)
PTC BROKEN
A 104
PT4 broken
(sterilization chamber wall)
PTC BROKEN
A 111
PT1 short-circuited
(sterilization chamber)
PTC SHORTCIRCUIT
A 112
PT2 short-circuited
(steam generator)
PTC SHORTCIRCUIT
A 113
PT3 short-circuited
(heating element)
PTC SHORTCIRCUIT
A 114
PT4 short-circuited
(sterilization chamber wall)
PTC SHORTCIRCUIT
A 200
Pre-heating not performed within the
timeout (heating resistor problem).
HEATING PROBLEM
Turning off the
equipment