Rev. 1
Card Gr1-
11
PT3 PROBE
430#00180000
Before servicing, switch off the equipment and unplug the power
supply cable from the mains socket
1. Remove the covers (see card Gr7-1 );
2. Remove the PT3 wiring from the CPU board;
3. Remove the cable marker no. 3 to be used on the new probe;
4. Access to the left side of the machine and withdraw the wiring up to the
thermal device;
5. Unlock the spring (1) of the chamber’s insulation layer to access the PT3
fixing;
6. Loosen the nuts (2) and withdraw the prove;
7. Shape the new probe as in figure and mount it, restore connections and
assembly all items proceeding in reverse order as above;
8. Switch on the unit;
9. Run a sterilization cycle.
cable marker
PT3 cable
PT3
1
2