Rev.1
Page 1-
47
Factory Data
(FACTORY DATA item on the SERVICE menu)
This option allows to re-enter the default data in the case of data lost or damaged memory.
Technical report
(TECHNICAL REPORT item on the SERVICE menu)
This option allows to obtain the data printout of the history register stored in the sterilizer memory.
PT1 correction
(PT1 CORRECTION item on the SERVICE menu)
This option allows to set the Ohm value of the internal probe replaced after a repair action. The following display is shown:
1500.7 Ohm +
PT1 CORRECTION
-
per confermare
Use key
+ / - to change the value displayed, push the key to confirm.