LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
PT3 BROKEN
Fault in the PT3 The temperature detected by
the PT3 is higher than 250°C
PT3 broken
Replace the PT3 - see card Gr1-10
PT3 uncalibrated.
Perform the calibration – see Attachment N
Check the temperature value on the display.
If = the room temperature:
- turn off and on the sterilizer more times;
- Replace PT3 – see card Gr1-10.
If differs for many °C:
- turn off and on the sterilizer more times;
- Replace PT3 – see card Gr1-10;
- Reset and calibrate the CPU board – see
Attachment N.
Check the calibration.
Perform the PT3 calibration– see Attachment M.
Complete sterilizer calibration – see Attachment N
Replace the CPU board – see card Gr1-1
Reset or failure on the data memory.
Reset and calibrate CPU board – see Attachment N
Replace the CPU board – see card Gr1-1