Page 3-
63
H 404
LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
Clean properly gasket and parabola board
PT1
FLUCTUATING
Temperature
fluctuating around the
threshold during the
process phase
Difference between PT1max
and PT1min greater than 2°C
Steam leakage from the door gasket
Replace the door gasket - see card Gr6-1
Identify the valve and clean it
Steam leakage from a valve
Replace the valve causing the problem
- see cards Gr2
Steam leakage from a pipe
Replace the pipe causing the problem
Problem in the water pump of the steam
generator
Replace the water pump - see card Gr3-1
EV6 valve does not open correctly
Replace EV6 - see card Gr3-1
Dismount and replace the O-ring in the steam generator
Problem in the steam generator
Replace the lower section of the steam generator
- see card Gr4-6
Fault in the board
Replace the CPU board - see card Gr1-1
PT1 failure
Replace PT1 - see card Gr1-8