Rev. 1
Page 2-
25
MANUAL CYCLE INTERRUPTION
The operator can manually interrupt the cycle at any time by pressing the START/STOP key for three seconds.
The command generates the error E999
, given that the cycle did not finish correctly. As a consequence, until safe conditions are reached, the
display shows, along a beep:
MANUAL STOP
LEVELLING
101.2 °C E999
-0.47 bar 26:01
When safe conditions are reached, the machine activates a special procedure
, first asking the user to manually unlock the door by displaying the
following instruction:
PRESS TO
UNLOCK THE DOOR
86.2°C E999
-0.02 bar 26:01
Press the key to unlock the door . The following message is then displayed:
MANUAL STOP
OPEN THE DOOR
85.8°C E999
-0.01 bar 26:01
Finally, when the door is opened, you will be asked to reset the device by the following message:
MANUAL STOP
RESET SYSTEM
85.5°C E999
-0.01 bar 26:01
> 3s