Rev.1
Page 1-
39
e If the last cycle completed correctly (or was interrupted by MANUAL STOP) it will be possible to reprint it in either NORMAL or
EXTENDED mode. If the last cycle was interrupted by an alarm (MANUAL STOP excluded) it only the EXTENDED mode will be
available.
Following the reprint command, this message will be displayed:
NOW PRINTING
THE REPORT
PLEASE WAIT...
which will remain on the screen until printing is finished.
Setting the tank Filling mode
(FILLING OPTIONS on the ADVANCED menu)
The internal tank can be filled either manually or automatically, in the latter case, drawing water from an external device (tank or demineralizer
Milldrop connected to the device- see Chapter, “Installation”).
After FILL OPTIONS is selected, the following menu appears:
→ AUTOMAT. FILLING +
MANUAL FILLING ↑
EXIT ↓
-
When AUTOMATIC FILL is selected, automatic filling is activated.
In this case, as reached the minimum water level (icon MIN on) in the internal tank, the equipment enable the auxiliary water feeding pump for a
pre-set time or a time needed to reach the maximum level (icon MAX on).
When the maximum level (MAX signal) is reached, the automatic system is deactivated.
e Only activate the automatic filling mode AFTER the external tank has been filled with high quality distilled water or the Milldrop has been
turned on. Also remember to open the tap on the external tank or the Milldrop.
When MANUAL FILL is selected, the internal tank must be filled manually (see the Chapter, “First Start-Up”).
Scroll through the items with the + and - keys; confirm with the ↵ key.