Rev. 1
Page 2-
6
After a brief interval, the display changes and shows the temperature and pressure values of the chamber, with the current date and time.
1 134°C POROUS
NORMAL DRYING
101.0 °C 30/08/02
0.01 bar 18:13:05
To cancel the selection, press ESC ↑ on the control panel.
e If no sterilization program is selected, the equipment cannot start a sterilization cycle, and the following message appears on the display,
with a beep:
SELECT A PROGRAM
PLEASE...
If you use a program that is inappropriate for the type of material to be sterilized (see Program summary table) the effectiveness of the
sterilization process is NOT GUARANTEED
.
RUNNING THE STERILIZATION PROGRAM
General
A sterilization cycle consists of a determined number of phases. The number and duration of the phases can differ for the programs, based on the
type of air extraction, sterilization process and drying method.
The electronic control system monitors the various phases, at the same time checking that the various parameters are respected; if any type of
anomaly is encountered during the cycle, the program is immediately interrupted, generating an alarm identified by a code, with a relative
message explaining the nature of the problem.
With this type of control, when you select a suitable sterilization program, you are guaranteed perfect sterilization under any conditions.