LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
Replace the water pump - see card Gr3-1
ATM3 TIMEOUT
3
rd
pressure rising
(from -.80 to 0.00 bar)
not performed within
the preset time
3
rd
pressure rising not
performed within 3’ timeout
The water pump of the steam generator
does not work properly
Replace the power board - see card Gr1-1
Clean the valve
Incorrect operation of the EV6 valve
Replace the EV6 valve - see card Gr3-1
Fault in the pipe connecting the steam
generator
Replace the pipe
Sterilization chamber overloaded
Advise the operator for properly loading of the sterilization
chamber.
Restore the safety thermostat and repeat the sterilization
cycle.
Intervention of the steam generator safety
thermostat
Replace the safety thermostat- see card Gr1-12
Replace the lower section of the steam generator
- see card Gr4-6
Steam generator clogged
Different distilled water must be used
Fault in the CPU board
Replace the CPU board - see card Gr1-1
CPU board uncalibrated
Send the sterilizer to the Mocom Service department
Check the continuity of the steam generator
cartridge
Replace the failed cartridge - see card Gr1-19