EXIT THE SETUP MODE
Completed the sterilizer configuration, proceed as follows to return in normal mode::
– Go to the first-level menu (see SETUP layout).
e To return to the first level from any current menu level, just select item EXIT of the current menu and confirm by ↵ key.
Alternatively, you can press ↑
(ESC) key one or more times.
– Select EXIT and confirm with the ↵ key. This text appears on the display:
MILLENNIUM B
SETUP COMPLETE
↵ to exit
to resume
After several seconds, the device returns to normal operation in STAND-BY mode.
DEFAULT SETTING
The sterilizer leaves the factory with the following settings:
DATE: current date
TIME: current time
PROGRAMMES: Preset 1: 134°C POROUS (standard drying)
Preset 2: 134°C HOLLOW (standard drying)
Preset 3: 134°C SOLID (standard drying)
Preset 4: 134°C EMERGENCY
e The programs indicated should be considered as preferential settings. However, other combinations are possible based on the
destination market.
ST-BY MODE: HIGH (pre-heating)
PRINT OPTIONS: OFF (1 copy)
FILLING OPTIONS: MANUAL
DRAIN OPTIONS: INTERNAL