Rev. 1
Page 3-
25
E 999
LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
MANUAL STOP
Manual interruption of
the cycle
START/STOP key pushed for
more than 3'' during the cycle
process
The user pushed the START/STOP key.
Advice the user to do not use the manual stop function when
you need to load an additional items to be sterilized. Suggest
to use the Emergency cycle.
Replace the keyboard - see card Gr1-3
The START/STOP key is blocked down.
Replace the CPU board - see card Gr1-1