A 040 (continue)
LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
Replace the fuse - see card Gr1-16
FILLING
PROBLEM
Fail of the automatic
filling from the
external tank.
The MlN level signaling is not
turned off within 2’ from the
start command of the
automatic filling.
Water pump not powered.
Replace the power board - see card Gr1-1
Break in the water pump wiring.
Replace the water pump wiring
Error in the water pump wiring connection. Restore the right connection of the water pump wiring.
Leak in the internal pipe.
Replace the broken pipe
Perforation in the main tank
Replace the main tank - see card Gr4-1
Disconnection of the level MIN float wiring. Restore the float wiring connection.
Fault in the level MIN float.
Replace the level MIN float - see card Gr4-2
Check the integrity of the pipe from the
Milldrop device
Restore the pipe.
Check the START/STOP cable connection
between sterilizer and MILLDROP.
Restore the connection.
Replace the START/STOP cable.
Check the MILLDROP operation.
Perform the maintenance on the Milldrop device –
See Operating Manual