Rev. 1
Page 3-
48
LCD INDICATION Alarm description Effect/Signaling
Possible cause / Check
Proposed solution
Identify the valve and clean it
Air leakage from a valve
Replace the valve - see cards Gr2
Air leakage from the heat exchanger
Replace the heat exchanger - see card Gr4-3
Air leakage from the pipe of the pressure
transducer.
Unplug and plug again the pipe correctly on the pressure
transducer, reset the alarm and restart the cycle
Pressure transducer uncalibrated
Send the sterilizer to the Mocom Service department
Check the type and the mass of the load
arranged in the chamber
Advice the user on loading and arranging the
material in the chamber as indicated on the
Operating Manual
Check the quality of the distilled water
Advice the user
Clean the hydraulic circuit
Replace the damaged components
Check the current arrangement of the
equipment
Change as necessary
Check the electric system of the room
Eliminate any possible power strip
The equipment must be powered through one’s own
supply line
Check the fan of the heat exchanger
Clean the fans – see card Gr4-4
Replace the damaged fan – see card Gr4-4