Rev.1
Page 1-
35
TIME OUT: XXX min
+/- to set
↵ to enter
to exit
It is possible to set a value between 0 and 300 minutes (in 30-minute increments), after which the heating elements are turned off (a condition
analogous to STAND-BY OFF), avoiding the useless consumption of electricity.
Set using the + and - keys; confirm with the ↵ key.
e This option is also active with STAND-BY OFF. However, in this condition the timer value obviously has no effect since the heating
elements are turned off anyway at the end of the sterilization program.
e When any cycle selection key (sterilization or test) is pressed, or the machine is turned off and on with the main switch, the original
STAND-BY mode (HIGH or LOW) is immediately reactivated.