CALCulate Subsystem (SPECM, K14) R&S FSL
1300.2519.12 6.398 E-11
C
haracteristics
RST value:
SCPI: device–specific
M
ode
SPECM
CALCulate<1|2>:MARKer<1...4>:SPECtrogram:Y:MAXimum[:PEAK]
This command positions the specified marker on the current maximum in y–axis direction
(constant x–axis value). The corresponding search in x direction is performed via the
CALCulate<1|2>:MARKer<1...4>:MAXimum[:PEAK] command.
This command is an event and therefore has no *RST value and no query.
The numeric suffixes <1|2> are irrelevant for this command.
Note: If no maximum value is found (level spacing to adjacent values < peak excursion), an
execution error (error code: –200) is produced.
Example
CALC:MARK:SPEC:Y:MAX
Positions marker 1 on the current maximum in y–axis direction.
Characteristics
RST value:
SCPI: device–specific
Mode
SPECM
CALCulate<1|2>:MARKer<1...4>:SPECtrogram:Y:MAXimum:ABOVe
This command positions the specified marker on the next maximum above its current position
(constant x–axis value). The corresponding search in x direction is performed via the
CALCulate<1|2>:MARKer<1...4>:MAXimum:LEFT and
CALCulate<1|2>:MARKer<1...4>:MAXimum:RIGHt commands.
This command is an event and therefore has no *RST value and no query.
The numeric suffixes <1|2> are irrelevant for this command.
Note: If no next smaller maximum value is found (level spacing to adjacent values < peak
excursion), an execution error (error code: –200) is produced.
Example
CALC:MARK2:SPEC:Y:MAX
Positions marker 2 on the current maximum in y–axis direction.
CALC:MARK2:SPEC:Y:MAX:ABOV