1xEV-DO BTS Analyzer (Option K84) R&S FSL
1300.2519.12 4.376 E-11
Restore Default Tables
Restores the predefined channel tables (see Predefined Channel Tables on page 4.303) to their
factory–set values. Existing channel tables with the same name as default channel tables are
replaced by this action. In this way, you can undo unintentional overwriting.
Remote: CONF:CDP:CTAB:REST
Result Settings
Opens the Result Settings dialog box.
The Result Settings dialog box contains the following parameters:
Select Code Slot
Channel Type
Mapping
Mapping Auto
Code Dom Overview
CDP Average
Code Power
Normalize
Select Code Slot
This softkey toggles between code selection and slot selection.
If the focus is on Code Selection, select a code number. Enter the code as a decimal. The range
depends on the specified channel type. For channel type PILOT and PREAMBLE values
between 0 and 31 are valid. For channel type MAC the range is between 0 and 63 and for DATA
only 63 is valid.
The following measurements take results for a Code into account: Power vs Symbol, Channel
Results, Bitstream, Symbol Constellation, EVM vs Symbol. In the following result displays, the
selcted code is highlighted in red:Code Domain Power, Channel Table, Code Domain Error.